Requirement 6

Must meet electrical safety standards IEC 60601

 RequirementSpecification
6Device must meet electrical safety standards IEC 60601.Type of material should be insulated (e.g. plastic)Leakage current cannot exceed 100 uA.
Justification The device is powered and will be in contact with the userMinimize potential risk from electric shocks.If material is not insulated, potential risk of shock increases. Leakage current limit is derived from sources & IEC 60601.

Verification: Measuring device leakage current to ensure it is under 100uA


Reference:

[1] International Electrotechnical Commission. “Medical electrical equipment – Part 1-1: General requirements for basic safety and essential performance.” Standard No. IEC 60601-1-1: 2020. IEC, 2020. [2] Livenson, A. R. (1992). Leakage currents in medical electrical devices. Biomedical Engineering, 26(3), 113-120