{"id":100,"date":"2023-09-22T01:54:46","date_gmt":"2023-09-22T01:54:46","guid":{"rendered":"https:\/\/engprojects.tcnj.edu\/chroma-scan-health\/?page_id=100"},"modified":"2023-11-17T21:43:46","modified_gmt":"2023-11-17T21:43:46","slug":"requirement-6","status":"publish","type":"page","link":"https:\/\/engprojects.tcnj.edu\/chroma-scan-health\/requirement-6\/","title":{"rendered":"Requirement 6"},"content":{"rendered":"\n<p><\/p>\n\n\n\n<h2 class=\"wp-block-heading\"><strong>Must meet electrical safety standards IEC 60601<\/strong><\/h2>\n\n\n\n<figure class=\"wp-block-table\"><table><tbody><tr><td>&nbsp;<\/td><td><strong>Requirement<\/strong><\/td><td><strong>Specification<\/strong><\/td><\/tr><tr><td><strong>6<\/strong><\/td><td>Device must meet electrical safety standards IEC 60601.<\/td><td>Type of material should be insulated (e.g. plastic)Leakage current cannot exceed 100 uA.<\/td><\/tr><tr><td><strong>Justification<\/strong>&nbsp;<\/td><td>The device is powered and will be in contact with the userMinimize potential risk from electric shocks.<\/td><td>If material is not insulated, potential risk of shock increases. Leakage current limit is derived from sources &amp; IEC 60601.<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p><strong>Verification:<\/strong> Measuring device leakage current to ensure it is under 100uA<\/p>\n\n\n\n<hr class=\"wp-block-separator has-alpha-channel-opacity\"\/>\n\n\n\n<p><strong>Reference:<\/strong><\/p>\n\n\n\n<p>[1] International Electrotechnical Commission. &#8220;Medical electrical equipment &#8211; Part 1-1: General requirements for basic safety and essential performance.&#8221; Standard No. IEC 60601-1-1: 2020. IEC, 2020. [2] Livenson, A. R. (1992). Leakage currents in medical electrical devices. Biomedical Engineering, 26(3), 113-120<\/p>\n","protected":false},"excerpt":{"rendered":"<p>Must meet electrical safety standards IEC 60601 &nbsp; Requirement Specification 6 Device must meet electrical safety standards IEC 60601. Type of material should be insulated (e.g. plastic)Leakage current cannot exceed 100 uA. Justification&nbsp; The device is powered and will be in contact with the userMinimize potential risk from electric shocks. If material is not insulated, &hellip; <a href=\"https:\/\/engprojects.tcnj.edu\/chroma-scan-health\/requirement-6\/\" class=\"more-link\">Continue reading <span class=\"screen-reader-text\">Requirement 6<\/span><\/a><\/p>\n","protected":false},"author":494,"featured_media":0,"parent":0,"menu_order":0,"comment_status":"closed","ping_status":"closed","template":"","meta":{"ngg_post_thumbnail":0,"footnotes":""},"class_list":["post-100","page","type-page","status-publish","hentry"],"_links":{"self":[{"href":"https:\/\/engprojects.tcnj.edu\/chroma-scan-health\/wp-json\/wp\/v2\/pages\/100","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/engprojects.tcnj.edu\/chroma-scan-health\/wp-json\/wp\/v2\/pages"}],"about":[{"href":"https:\/\/engprojects.tcnj.edu\/chroma-scan-health\/wp-json\/wp\/v2\/types\/page"}],"author":[{"embeddable":true,"href":"https:\/\/engprojects.tcnj.edu\/chroma-scan-health\/wp-json\/wp\/v2\/users\/494"}],"replies":[{"embeddable":true,"href":"https:\/\/engprojects.tcnj.edu\/chroma-scan-health\/wp-json\/wp\/v2\/comments?post=100"}],"version-history":[{"count":0,"href":"https:\/\/engprojects.tcnj.edu\/chroma-scan-health\/wp-json\/wp\/v2\/pages\/100\/revisions"}],"wp:attachment":[{"href":"https:\/\/engprojects.tcnj.edu\/chroma-scan-health\/wp-json\/wp\/v2\/media?parent=100"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}