Device Requimrents & Specifications

  1. The device must be able to detect inflammation presented on the skin surface for all different skin tones

Justification:-

Skin discoloration is a symptom of reduced blood flow, which is a correlate for DVT. 

Specifications:-

1.1) The device correctly identifies  presentation of erythema from digital images regardless of skin tone over 85% of the time.3

Verifications:-

1.1) Test the algorithm using a new dataset of skin images and verify that 85% of true positives are identified.

  1. The device must be able to obtain the patient’s Hemoglobin bound Oxygen Levels.

Justification:- 

Systemic O2 levels are related to blood clot formation.

Hemoglobin bound Oxygen accounts for 97% of Systemic Oxygen.

Specifications:-

2.1) Obtains measurement at 1Hz from hospital oximeter.5

Verifications:-

2.1) Find the frequency at which the oximeter is taking measurements to ensure it’s no less than 1Hz.

  1. The device must accommodate a wide range of sizes to match the user.

Justification:- 

A wide range of sizes will allow the device to be worn comfortably as well as prevent the device interfering with DVT assessment.

Specifications:-

3.1) Ankle (Mean ± SD)

Male: 23.5 cm ± 1.5 cm (19.0 – 28.0 cm) 

Female: 21.9 cm ± 1.3 cm (18.0 – 25.8 cm)

3.2) Must be able to adjust circumference to accommodate the location’s size

Verifications:-

3.1 & 3.2) Verify that the size of the device will be incrementally changeable to accommodate the desired ranges by measuring the device’s length’s. 

  1. The device must have sufficient battery life to monitor the patient’s stats for long periods of time.

Justification:- 

A long battery life will allow for continuous monitoring of patients without disruption.

Specifications:-

4.1) More than 12 hours to allow changing of nurse shifts.

Verifications:-

4.1) Measure power consumption of the device and its constituent parts using a LabView model and/or test the physical parts specifications.

  1. Device must cause little to no reaction to the skin of the patient.

Justification:- 

Reactions to the skin can cause interferences with monitoring results and can change patient DVT detection.

Specifications:-

5.1) The power delivered to the skin from the LEDs is less than 180 mW/cm2 at a rate of 0.2 mJ/cm2.6

5.2) The material used that contacts skin should be compliant with ISO 10993-10:2010(E)7.

Verifications:-

5.1) Measure the power output from the LED’s dedicated device to ensure it remains within desired range. 

5.2) Place material on participant skins and check for discoloration, damage, and inflammation. 

  1. The device must be able to alert healthcare workers when DVT risk reaches threshold level set by healthcare workers.

Justification:- 

Alerting healthcare workers will allow for preventive measures to be taken and prevent further clot formation and damage to the blood vessels.

Specifications:-

6.1) Trigger alarm within <1s when the threshold is reached.

6.2) SpO2 threshold can be set between 95% and 98% with 1% resolution.

6.3) Erythema threshold can be set between 10-25% of the area of the skin displaying inflammation with 1% resolution.

Verifications:-

6.1) Time the delay between the artificial triggering of the device and the triggering of the alarm. 

6.2) Test that the device SpO2 threshold can be altered by a user within the given ranges and resolution.

6.3) Test that the device erythema threshold can be altered by a user within the given ranges and resolution.

  1. The device must remain comfortable to wear.

Justification:- 

User comfortability is important as the user will be wearing the device for long periods of times, at least 12 hours. 

Leave a Reply

Your email address will not be published. Required fields are marked *