Requirement: The device must reduce the pressure at the device-user interface during proper assisted ambulation relative to axillary crutches
Methods: Users ambulated with the Ergo-Crutch for 20 strides while pressures were recorded.
Acceptance Criteria: Total maximum pressure must be less than 110 kPa.
Results: All maximum pressures were less than 110 kPa.

Outcome: Pass