Electronic System

Requirement: The device must monitor usage-metrics and provide feedback to the user.

Methods: Users with varying masses ambulated using the device. LED activity and pressure data were recorded.

Acceptance Criteria: LEDs turn on when recorded pressures are greater than 60.0 kPa (30.0 kPa per crutch) and the electronic systems on the device record user pressure metrics.

Results: The LEDs turned on when the recorded pressures were greater than 60.0 kPa (30.0 kPa per crutch) and all pressure data was recorded by the electronic systems on the device.

Outcome: Pass