Specification A: The device should have the sensing capacity to monitor pressures up to 110 ± 1 kPa.
Methods: Increasing loads were applied to the load cell test bench.

Acceptance Criteria: The electronic system accurately measures the applied pressures up to 65 kPa ± 1 kPa.
Results: No significant difference between expected and measured pressures (p>0.05)
Outcome: Pass