Specification B: The device should provide a visual alert in the user’s line of vision when the total pressure applied to the device exceeds 75% of the maximum pressure.
Methods: Increasing loads were applied to the load cell test bench and LED activity was observed.

Acceptance Criteria: LED turns on when threshold pressure is reached.
Results: LED turned on when threshold pressure was exceeded, and turned off when pressure was lifted.
Outcome: Pass