{"id":18,"date":"2019-09-05T14:32:30","date_gmt":"2019-09-05T14:32:30","guid":{"rendered":"http:\/\/engprojects.tcnj.edu\/heatedmassage\/?page_id=18"},"modified":"2019-11-24T01:05:50","modified_gmt":"2019-11-24T01:05:50","slug":"design","status":"publish","type":"page","link":"https:\/\/engprojects.tcnj.edu\/heatedmassage\/design\/","title":{"rendered":"Design"},"content":{"rendered":"\n<p>Final Design Requirements and Specs<\/p>\n\n\n\n<table class=\"wp-block-table has-fixed-layout is-style-stripes\"><tbody><tr><td><strong>Requirements&nbsp;<\/strong><\/td><td><strong>Specifications<\/strong><\/td><td><strong>Justification<\/strong><\/td><\/tr><tr><td>Device must apply heat to    the intended tissue.&nbsp;<\/td><td>Device must reach <br>41+\/- 2 \u00b0C.<br><br><br><\/td><td><br>Pennes HH: Analysis of tissue and arterial blood temperatures in the resting human forearm: 1948. J Appl Physiol. 1998, 85 (1): 5-34.<br><a href=\"https:\/\/diagnosticpathology.biomedcentral.com\/articles\/10.1186\/1746-1596-9-36\">https:\/\/diagnosticpathology.biomedcentral.com\/articles\/10.1186\/1746-1596-9-36<\/a><\/td><\/tr><tr><td>Device must securely fit a wide  range of patients<\/td><td>1a.Device must be adjustable between a forearm circumference of 15-38 centimeters<br>1b.Device should not move by 1cm after 10 min<\/td><td>Its +3 sigma for men and -3 sigma for women<br>Device must fit 6 sigma of human forearms&nbsp;<br>AF AMRL-TR-180-119 McConville, John Anthropometric Relationships of Body and Body Segments Moments of Inertia Anthropology Research Project, Inc. A\/F Aerospace Medical Res 12-80<\/td><\/tr><tr><td>Device must be portable\/            Device must be light enough to carry without causing strain to  the user<\/td><td>It will fit in an average size backpack of (16.7&#8243; x 13&#8243; x 8.5&#8243;).\u00a0\u00a0<br><br>Device will be under 8 pounds<br><\/td><td><br>  <br>&#8212;  average volume size of a          backpack&nbsp;<br>&nbsp;<\/td><\/tr><tr><td>Device will have enough power to support multiple uses per     day.&nbsp;<\/td><td>Device must last for at least two hours of constant use (6 times = 1 hours)\u00a0<\/td><td><br><br><br><\/td><\/tr><tr><td>Device should function for a       pre-sent amount of time<\/td><td>Device will run for 10 minutes<\/td><td><br><\/td><\/tr><tr><td>Device will have a manual            shutoff.<\/td><td>On and off button located on      the surface of the device for      easy access to the user<\/td><td>If patient feels <br>uncomfortable or pain theycan shut it off manually<\/td><\/tr><tr><td>Device will have an automatic    fail-safe shut off <\/td><td>Device will automatically shut   off when Device temperature    >44\u00b0C.<br><br><\/td><td>To ensure no tissue damage occurs.Dewhirst, Mark W et al. \u201cTHERMAL DOSE REQUIREMENT FOR TISSUE EFFECT: EXPERIMENTAL AND CLINICAL FINDINGS.\u201d <em>Proceedings of SPIE&#8211;the International Society for Optical Engineering<\/em>  vol. 4954 (2003): 37. doi:10.1117\/12.476637<\/td><\/tr><tr><td>Device will apply soft tissue      massage<br><br><\/td><td>The pressure range will be from 2 PSI to 4 PSI<br>Roberts, Langdon. \u201cEffects of    patterns of pressure application on resting electromyography during massage.\u201d <em>International journal of therapeutic massage &amp; bodywork<\/em> vol. 4,1 4-11. 30 Mar. 2011&nbsp;<br><br><\/td><td>Soft tissue massage increases   blood flow to a particular            region<sup>1<\/sup><br>Soft tissue massage can               promote healing via increased  fibroblast recruitment<sup>1<\/sup>&nbsp;<br><br><br><\/td><\/tr><tr><td>Device should be rechargeable<\/td><td><br><\/td><td><br><\/td><\/tr><\/tbody><\/table>\n","protected":false},"excerpt":{"rendered":"<p>Final Design Requirements and Specs Requirements&nbsp; Specifications Justification Device must apply heat to the intended tissue.&nbsp; Device must reach 41+\/- 2 \u00b0C. Pennes HH: Analysis of tissue and arterial blood temperatures in the resting human forearm: 1948. J Appl Physiol. 1998, 85 (1): 5-34.https:\/\/diagnosticpathology.biomedcentral.com\/articles\/10.1186\/1746-1596-9-36 Device must securely fit a wide range of patients 1a.Device must &hellip; <\/p>\n<p class=\"link-more\"><a href=\"https:\/\/engprojects.tcnj.edu\/heatedmassage\/design\/\" class=\"more-link\">Continue reading<span class=\"screen-reader-text\"> &#8220;Design&#8221;<\/span><\/a><\/p>\n","protected":false},"author":220,"featured_media":0,"parent":0,"menu_order":0,"comment_status":"closed","ping_status":"closed","template":"","meta":{"ngg_post_thumbnail":0,"footnotes":""},"class_list":["post-18","page","type-page","status-publish","hentry"],"_links":{"self":[{"href":"https:\/\/engprojects.tcnj.edu\/heatedmassage\/wp-json\/wp\/v2\/pages\/18","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/engprojects.tcnj.edu\/heatedmassage\/wp-json\/wp\/v2\/pages"}],"about":[{"href":"https:\/\/engprojects.tcnj.edu\/heatedmassage\/wp-json\/wp\/v2\/types\/page"}],"author":[{"embeddable":true,"href":"https:\/\/engprojects.tcnj.edu\/heatedmassage\/wp-json\/wp\/v2\/users\/220"}],"replies":[{"embeddable":true,"href":"https:\/\/engprojects.tcnj.edu\/heatedmassage\/wp-json\/wp\/v2\/comments?post=18"}],"version-history":[{"count":0,"href":"https:\/\/engprojects.tcnj.edu\/heatedmassage\/wp-json\/wp\/v2\/pages\/18\/revisions"}],"wp:attachment":[{"href":"https:\/\/engprojects.tcnj.edu\/heatedmassage\/wp-json\/wp\/v2\/media?parent=18"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}