Verification

Device RequirementVerification Activity
1. Device will operate under a maximum angular velocity for the knee joint corresponding to specific gait velocitiesGoniometer will be used to measure positions at each angle. Video recording will record the angular position over time. Angular velocity will be calculated from these values.
2.1 Device must provide resistance during active flexion

2.2 Device must provide assistance during active extension
Mathematical model of the force vector at the point of insertion below the knee to resolve torque
3. Device must allow normative knee joint angles for the four phases of walking gaitUse a goniometer to measure knee angles at maximum extension and flexion
4. Must accommodate the middle 90% of adults 60 years old and upMeasure the maximum and minimum distance between the top and bottom sections of the device using a flexible tape measure. Measure the maximum and minimum circumference at each section with straps using a flexible tape measure.