Design Inputs

RequirementJustificationSpecificationJustification

1. The device must stay secure for a clinically acceptable period.
It is necessary to ensure the device does not leak and adapts to daily usage.
Must be able to withstand a displacement force of 25.51 ± 17.90 lbs for 24 hours.
The attachment site component needs to be changed every 24 hours for sanitary purposes while accommodating daily usage.
2. The Device Must Collect Urine Outputs.Minimizes patient cleaning efforts for nursing staff, and ensures that patients are staying dry, and urine outputs are accurate.Can collect a flow rate of 4.5-21.6 mL/sec for up to a minute.
Uroflowmetry indicates these values to be the normal ranges for a urination cycle among males.
3. Can accommodate 95% of males.The device can be used for at least 95% of male patients who would benefit.Can accommodate 6-16.5 cm in length, and 7.5-13.5 cm in girth.The range includes 95% of penis sizes; 6-12.5cm in length and 7.5-11 cm girth for flaccid, and 10-16.5cm in length and 9.5-13.75 cm girth erect.
4. Clinical staff must be able to easily collect a clean urine sample from the canister.Allows urine samples to be obtained for urinalysis.The urine collection point is sterile (SAL 10-6), and the device provides a non-mixed sterile sample that was voided less than 1 hour ago.Prevents urine sample contamination via the device, and provides an adequate sample for urinalysis (mixed samples and samples over 1 hour old are not used for urinalysis unless refrigerated).
5. Must provide a warning to clinical staff when the canister is full.Allows clinicians to empty the canister before a leakage event since urine is considered a biohazard.Detect volume between 0 – 2000 ml ± 50 ml at any arbitrary time, then trigger an alarm to alert staff.2000ml is the full canister volume and an overflowing urine canister is a biohazard.
6. The device must be able to regulate the negative pressure based on the urination cycle.Necessary to reduce vac-related injuries and improve patient/user comfort.Turn on negative pressure from 40-80 mmHg ±5 mmHg using a urination sensor.Current vac-assists operated 40-80 mmHg.
7. The device must ensure proper house vacuum settings.Necessary that clinical staff does not set vac to outside the accepted range.
Detect negative pressure set to >80 mmHg or <40 mmHg ± 5.Negative pressure >80 mmHg or <40 mmHg is outside the acceptable range.
RequirementSpecification
8. The collected volume must be visible to clinical staff.Contains measuring lines per every 50 mL or digital display of volume ± 50 ml.
9. The canister must hold normal urine volume output during the clinically relevant time.2000 ml to accommodate the upper range.
10. Must pass electrical safety standards for hospital-based systems.500 microamps for leakage current, and less than 0.1 ohms for ground resistance (IEC 6060).
11. The device captures 90% of moisture passing to the vacuum.The device captures 90% of moisture passing to the vacuum[13].