Verification & Validation


RequirementConstraintVerificationValidation
1. The device must allow adjustment of plantarflexion and dorsiflexion to match the patient’s current phase of rehabilitation.– The device must be easy/intuitive to adjust ROM settings.
– The device must minimize the risk of tissue damage, falls, or excessive fatigue.
– The device must minimize the risk of reinjury.
TBDTBD
2. The device must allow adjustment of abduction and adduction to match the patient’s current phase of rehabilitation.– The device must be easy/intuitive to adjust ROM settings.
– The device must minimize the risk of tissue damage, falls, or excessive fatigue.
– The device must minimize the risk of reinjury.
TBDTBD
3. The device must allow adjustment of inversion and eversion to match the patient’s current phase of rehabilitation.– The device must be easy/intuitive to adjust ROM settings.
– The device must minimize the risk of tissue damage, falls, or excessive fatigue.
– The device must minimize the risk of reinjury.
TBDTBD
4. The device must actively assist the ankle range of motion during the active recovery phase.N/ATBDTBD
5. The device must tolerate loads without loss of structural integrity.– The brace must be durable enough to withstand daily long-term use.TBDTBD
6. The device must be lightweight.– The brace must be designed with inclusivity of a wide range of age groups and mobility levels.
– The device must minimize the risk of tissue damage, falls, or excessive fatigue.
– The device must minimize the risk of reinjury.
TBDTBD
7. The device must be comfortable.– The brace must be designed with inclusivity of a wide range of age groups and mobility levels.TBDTBD
8. The device must allow fast switching between the three rehabilitation phase modes.N/A– The device’s mode control will be confirmed by testing the success rate and duration of the mode switching.TBD
9. The device must constrain the range of motion of the ankle to match the patient’s current phase of rehabilitation.N/A– The device’s accuracy will be confirmed by comparing range of motion readings to a fixture with set angles.TBD
10. The device must provide a controllable active element capable of assisting the patient through a prescribed range of motion during the sub-acute phase.N/A– The device’s active component control will be confirmed by comparing range of motion readings to a fixture with set angles.TBD
11. The device must allow clinicians to set a custom range of motion boundary for each plane of motion within the sub-acute phase.N/A– The device’s active component control will be confirmed by comparing range of motion readings to a fixture with set angles.TBD
12. The device must include a fail-safe that prevents motion beyond clinician-set limits.N/A– The device’s fail-safe mechanism will be confirmed by attempting to rotate the device beyond the constrained range of motion and measuring further motion beyond that range.TBD
13. The device must include an emergency shut-off mechanism that rapidly stops all powered motion.N/A– The function of the shutoff control will be confirmed by the the measurement of time that it takes for the device to return to a neutral position from the time that the shut off was activated.TBD