Verification Testing
Below are the anticipated verification studies.
- Comfortability
- Design Input Addressed: The device must be comfortable for the patient.
- Objective: Utilize a Shore hardness and elastic modulus test to confirm the softness of the material.
- Verification Activity: Conduct Instron testing to calculate elastic modulus and a Shore Hardness test to compare to literature values.
- Volume Verification
- Design Input Addressed: The device must collect an acceptable urine sample volume.
- Objective: Confirm that the device collects and transports a urine sample of at least 2.0 mL, the minimum required for urinalysis.
- Verification Activity: Run a load cell to detect a sample volume of 2.0 ± 0.5 mL
- Biomarker Detection Accuracy
- Design Input Addressed: The device must be able to detect the presence of bioburden associated with UTIs.
- Objective: Verify the device correctly identifies nitrite concentrations at or above 2 mg/L as positive and below 2 mg/L as negative, and leukocyte concentrations at or above 10 leukocytes/HPF as positive and below 10 leukocytes/HPF as negative.
- Verification Activity: Using prepared urine samples with known nitrite and leukocyte concentrations, a colorimetric sensor measures dye intensity resulting from the Griess reaction (nitrites) and a two-step chemical reaction (leukocyte esterase), with sensor readings compared to spectrophotometer results.
- Result Communication
- Design Input Addressed: The device must communicate the levels of bioburden detected in the urine sample.
- Objective: Ensure that the display correctly classifies urine test results as negative, caution, or positive based on predefined biomarker thresholds.
- Verification Activity: Generate test samples with varying biomarker levels and verify that the display output matches expected results.
- Electrical Safety
- Design Input Addressed: The device must meet electrical safety standards IEC 60601-1.
- Objective: Ensure that the device’s leakage current remains below 500 μA under normal operating conditions.
- Verification Activity: Run the device and measure leakage current using a digital multimeter to confirm compliance with IEC 60601-1 standards.
- Sanitization Effectiveness
- Design Input Addressed: The device must be sanitizable.
- Objective: Validate that EPA-approved disinfectants effectively sanitize the device, preventing false positives and contamination.
- Verification Activity: Run a cleaning sequence after a positive sample, then test with a negative sample to confirm no residual contamination.