Design Inputs & Considerations

        Breif Req         Requirement              Justification       Specifcation
     Attachment  The device must stay attached to the user throughout its use.  The device is to be worn  on the legs during its  use, so it must stay  attached so the device can function as intended. 1.1 The device must not be displaced 5% of the total lower leg length after use.
             Size  The device must be able to fit into a confined area. Places of intended use are typically of small dimensions such as office space, truckers, or passengers on
a plane
2.1 The device attached to patient’s leg
must function within a seat pitch (distance
from one point on a seat to that same
point of the seat behind it) of 28 in.
2.2 The device must allow at least a 120
deg angle of the lower leg to ensure
blood flow is not 100% constricted.
               Fit  The device must be able to fit the
average circumference of a male and
female human leg or foot when in use.
VTE effects both males and females equally. Recurrent VTE is higher in males and pregnancy increases risk for females. 3.1 The device must fit average male and
female circumference of the upper calf of
35.9 ± 2.4 cm and 34± 2.5 cm respectively
3.2 The device must fit average male and
female circumference of the ankle of
23.4 ± 1.5 cm and 21.9 ± 1.3 cm
respectively.
      Compression  The device must apply intermittent
compression to the human lower leg.
Compression mimics
calf muscle contraction
in order to increase
blood flow to the
popliteal vein. Intermittent pneumatic
compression devices
are currently the best
modality of prevention of
VTE.
4.1 The compression must apply between
80 and 120 mmHg of pressure to the lower
leg.
4.2 When activated by the sensors
the compression device must apply
pressure for 50 seconds followed by 50
seconds of deflation. This must continue
until the sensors deactivate the
compression.
              Sound The device must not produce
obtrusive sounds and noises.
Reduce social stigma of wearing the
device and ensure that others aren’t
bothered by the noise produced.
5.1 The level of noise that the device
produces must not exceed 55 dB.
    Control System  The device must activate and inactivate
due to specific threshold measurements
from sensors.
Individuals are only at risk for VTE if the
correlates are detected above a specified
threshold.
6.1 The device must activate the
compression system once a 1℃ increase
is detected by the temperature sensor.
6.2 The device must activate the
compression system once a 100 mL
increase in lower leg volume is detected.
6.3 The device must deactivate the
compression once the temperature and
volume sensors return to their baseline
measurements.
           Battery
 (Duration of Use) 
The device’s battery must last for the duration of at risk
period for VTE.
Intended use is during long
haul flights of 12 hours, an 8 hour work day
and periods of sitting +4 hrs. The device
must be able to stay powered for monitoring
and during periods of activationthroughout
the whole sitting.
7.1 The device’s battery must be able to
provide 12 hours of functionality including
monitoring and compression.
    Comfortability  The device must be made
out of a comfortable material
The device will be worn for hours at a time, so the
device must be comfortable where the user won’t be irritated
when wearing the device.
8.1 The device must score XXX on a survey.
      TSA Approved  The device must pass
TSA guidelines for flights
An intended user is those
on long haul airplane
flights, so the device must be allowed to go on an airplane.
9.1 The device must pass TSA prescreening.
   Leakage Current Leakage current can lead to an
electric shock, which is
dangerous to the user.
Device must comply with
IEC 60601-1 series standards
regarding battery operated
medical devices.
10.1 IEC 60601-1 series.
 Skin Temperature      Measurement  The device must measure local
skin temperature in front of the tibia
According to one study,
skin temperature in the
front of the tibia increased
significantly during a study
in which subjects remained
sitting for 12 hours. Increases were
seen in 4 – 12 hours.
11.1 Device must be able to measure skin
temperature of 32℃ +/- 2 degrees
11.2 Must be able to differentiate within
a 1°C increase.
 Volume  Measurement  The device must measure volume
in the leg below the knee.
According to one study,
the leg volume in the calf
increased significantly
during a study in which
subjects remained sitting
for 12 hours. Increases
were seen in 4 – 12 hours.
12.1 The device must detect volume
increases of 4%.
12.2 The device must be able to be
calibrated to the user’s normal leg size.