Project

Device Requirements:

Design RequirementDesign SpecificationVerificationValidation
1. The device will
restrict motion of
the arm to the
scapular plane
(defined as 30-45⁰
anterior to coronal
plane).
1.1. The scapular
plane component
of the device will
be 37.5⁰+1⁰, -7.5⁰
anterior to coronal
plane.
1.1. Ten measurements of
the angle between the
scapular plane
component of the device
and the coronal plane will
be measured using a
goniometer. The average
and standard deviation
will be calculated. For
acceptance, ±3σ must fall
within the specifications.
1.1. Ten measurements
of the angle between the
center of the test
subject's arm and the
coronal plane will be
measured using a
goniometer. The average
and standard deviation
will be calculated. For
acceptance, ±3σ must
fall within the
specifications.
2. The device will
move in external
rotation.
2.1. The arm must
be abducted 90⁰ ±
10⁰ in the scapular
plane.
2.2. External
rotation must be
performed in a
range of (90⁰-
180⁰) ± 10⁰ .
2.1. On the device, the
angle of abduction will
be set to the
specification. Ten
measurements of the
angle will be measured
using a goniometer. The
average and standard
deviations will be
calculated. For
acceptance, ±3σ must fall
within the specifications.
2.2. On the device, the
angle of external rotation
will be set to the
minimum and maximum
specification range values
and measured using a
goniometer at each angle
using the same
parameters in 2.1.
2.1. On the device, the
angle of abduction will
be set to the
specification. Ten
measurements of the
angle between the test
subject's side and center
of abducted arm will be
measured using a
goniometer. The average
and standard deviations
will be calculated. For
acceptance, ±3σ must
fall within the
specifications.
2.2. The test subject will
move the device from
the minimum to
maximum allowable
angles within these
ROM and these angles
will be measured using a
goniometer with the
same parameters in 2.1.
3. The device will
move in
abduction.
3.1. Abduction
must be performed
in range of motion
of (0⁰-120⁰) ± 10⁰.
3.1. On the device, the
angle of abduction will
be set to the minimum
and maximum
specification range values
and measured using a
goniometer at each angle.
The average and standard
deviations will be
calculated. For
acceptance, ±3σ must fall
within the specifications.
3.1. The test subject will
move the device from
the minimum to
maximum allowable
angles within this ROM
and these angles will be
measured using a
goniometer with the
same parameters in 2.1.
4. The device will
provide variable
resistance training.
4.1. Resistance
will vary from 2.0
± 0.5 lb to 7.0 ±
0.5 lb.
4.1. Stiffness will be
experimentally
determined for each
resistive component
according to the k value
for resistance
requirements of Theraband
resistance bands at
100% elongation. A
variation of loads will be
applied to the resistive
component. The change
in length will be
measured and compared
against the force
produced by the load.
This will generate a plot
and the slope (or
stiffness) will be
calculated. The average
and standard deviation
for the stiffness of each
resistive component will
be calculated and ±3σ
must fall within the
specifications for
acceptance criteria.
N/A
5. The device
must fit the
average range of
human sizes.
5.1. The device must fit
the average length of a
shoulder (13.7 ± 1.3
cm).
5.2. The device must fit
the average torso
(circumference of 57.2
cm - 99.1 cm) ± 1.0 cm.
5.3. The device must fit
the average wrist.
(circumference of 14.0
cm - 22.6 cm) ± 1.0 cm.
5.4. The device must fit
the average length of the
brachium (32.4 - 43.1
cm) ± 1.0 cm.
5.5. The device must fit
the average length of the
anti-brachium (24.6 cm -
26.4 cm) ± 1.0 cm.
Note: Ranges for 5.2-5.5
due to adjustable
property of component.
5.1, 4, 5. Ten
measurements will be
obtained for the
length of the
shoulder, brachium
and anti-brachium
components of the
device using a ruler.
The average and
standard deviation
will be calculated.
For acceptance, ±3σ
must fall within the
specifications.
5.2-3. Same
parameter as above
will be used for
measurements of
wrist and torso using
a tape measure.
5.1-5. The device will
be tested on multiple
subjects and the range
of fit will be
examined. The ranges
should fit 90% of the
human subjects.
6. The device
must be
comfortable.
6.1. The device must be
worn for up to 30
minutes.
N/A6.1. Test subjects will
wear the device for at
least 30 minutes and
comfort will be
ranked from 1 (very
uncomfortable) to 5
(very comfortable)
7. The device will
not significantly
increase exercise
time.
7.1. The device will not
increase exercise time
by more than 10%.
N/A7.1. The time it takes
to don and doff the
device will be
recorded per test
subject.
8. The system
actively records the
degree of motion the
user achieves.
8.1. A sensor must
measure angles from
(0⁰ - 180⁰) ± 10⁰.
8.1 On the device, the
angles of abduction
and internal rotation
will be set to the
minimum and
maximum angle
values and measured.
Ten measurements
will be taken for each
angle both using a
goniometer and from
the output of the
device. A paired t-test
will be used to
determine of the if
the values are
statistically the same.
N/A
9. The system must
have a portable
power source.
9.1 A 9V battery
powers the system.
N/A9.1. Device will be
checked to determine
if a 9V battery is
present.
10. The system
provides a real-time,
visual representation
of the degree of
motion.
10.1. A wireless
module transmits data
between the
microcontroller and a
smartphone.
10.2 An application
on the smartphone
plots angle
measurements in real
time.
N/A10.1-2. The test
subject will perform
the specified
exercises and the
smartphone will be
examined to see if
data is being plotted
in real time.
11. The system must
store the average
effort per trial.
11.1 The application
on the smartphone
must overwrite the
time, date, and
average effort per
trial.
N/A11.1. The test subject
will perform two sets
of the specified
exercise and the
smartphone will be
observed to examine
if the average effort
is overwritten for
second trial.

Justifications:

1.1. The scapular plane of motion is used for rehabilitation because it facilitates optimal muscle forces. Excessive motion in the coronal plane may cause overload on the anterior cuff tendons. There is also a greater external rotatory force and smaller compressive force in this plane.

2.1-2. Values obtained from the International Journal of Sports Physical Therapy.

3.1. Values obtained from appropriate literature.

4.1-2. Justified during testing.

5.1-5: Values obtained from appropriate literature.

6.1. Typical physical therapy sessions about 30 minutes.

7.1. Justified during testing.

8.1. Recommended by TCNJ Trainer, Megan Guicheteau.

9.1. Most cost effective solution.

10.1-2. The user can easily view their progress.

11.1. The user can monitor day-to-day progress.